Analysis range |
0~100% |
|
|
Resolution ratio |
1% |
|
|
Error |
Bottle <3%; water sample <3% |
|
|
Repetitiveness |
10% |
|
|
Range drift |
±5% F.S. |
|
|
Pure water light loss |
±3% |
|
|
Determination period |
Within 1 to 24 hours, it is set in hours |
|
|
Analysis mode |
integral point |
Run by the hour, up to 24 groups per day |
|
margin |
Run for 0-9999 minutes at intervals |
|
|
hand movement |
Manual analysis |
|
|
long-range |
Remote control by MODBUS; dry contact control |
|
|
Calibration method |
Number of intervals |
The interval was calibrated according to the cumulative analysis |
|
Number of days between intervals |
Interval calibration is performed according to the number of running days |
|
|
hand movement |
Manual calibration |
|
|
long-range |
Remote control is performed in MODBUS mode |
|
|
Cleaning method |
voluntarily |
Analysis cleaning; fault cleaning; abnormal power failure cleaning; emergency shutdown cleaning |
|
hand movement |
Clean by hand |
|
|
report to the police |
Low level alarm; self-test alarm; fault alarm; out of specification alarm (one relay) |
|
|
Communication |
An RS232 or RS485 interface standard MODBUS protocol |
|
|
Memory |
About 20,000 groups (power off automatic save); support USB drive export EXCEL; |
|
|
Put a seal on |
Micro industrial thermal printer (optional) |
|
|
Maintenance cycle |
About 30 days, 45 minutes |
|
|
Cell display |
7 inch LCD, 800*480 resolution, TFT true color touch screen |
|
|
Service environment |
5-35℃, humidity <90% (no condensation) |
|
|
Water sample temperature |
5—35℃ |
|
|
Water harvesting unit |
Used for water sample circulation and retention; automatic backwash precision filtration system (optional) |
|
|
Size |
90cm×65cm×43cm |
|
|
Weight |
55KG |
|
|
Source |
AC 220V ± 10%, 50Hz ± 1% |
|
|
Power |
150W |
|